Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-55
Select Output Range
The Select Output Range screen is used to select the hardware range for the
selected analog output channel. Possible ranges for the voltage outputs are:
0-100 mV, 0-1, 0-5, 0-10 V.
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Select Channel >
Select Range
.
•
Use
and
to move the cursor up and down.
•
Press
to save the new range.
OUTPUT CHANNELS:
>ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
ª
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
RANGE AVG DIAGS ALARM
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-5V ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM