C-Link Protocol Commands
Communications Configuration
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-35
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example that follows saves the parameters to FLASH
memory.
Send:
set save params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example that
follows reports that the internal time is 2:15:30 pm.
Send:
time
Receive:
time 14:15:30
set time
hh:mm:ss
hh
= hours
mm
= minutes
ss
= seconds
This command sets the internal clock (24-hour time). The example that
follows sets the internal time to 2:15 pm.
Note
If seconds are omitted, the seconds default to 00.
Send:
set time 14:15
Receive:
set time 14:15 ok
Communications
Configuration
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr dns
Receive:
addr dns 192.168.1.1
set addr dns
address
This command sets the domain name server
address
, where
address
consists
of four numbers ranging from 0-255 inclusive, separated by “.”.
Send:
set addr dns 192.168.1.1
Receive:
set addr dns 192.168.1.1 ok
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...