Servicing
Ozonator Lamp Heater Replacement (Optional)
7-40
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Ozonator Lamp Heater
Replacement
(Optional)
Use the following procedure to replace the ozonator lamp heater.
Equipment Required:
Ozonator lamp heater
Allen wrenches: 3/32-inch, 7/64-inch
Philips screwdriver
Heat conductive compound
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Unplug heater cable from the LAMP HEATER and LAMP connectors
on the ozonator board (
3. Disconnect plumbing from the ozonator.
4. Remove the ozonator by removing the two screws securing the ozonator
to the floor plate and lifting the ozonator straight up.
5. Remove the ozonator flange by removing three screws securing the
bottom flange and pulling the flange off.
6. Remove the ozonator lamp heater block from the ozonator by removing
four Allen screws.
7. Coat new lamp heater block with thin film of heat conductive
compound.
8. Install new ozonator lamp heater block by following the above
procedure in reverse.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...