Thermo Fisher Scientific
Model 49
i
Instruction Manual
4-1
Chapter 4
Calibration
This chapter describes how to perform a multipoint calibration of the
photometric ozone analyzer. It is based upon the current EPA approved
procedure using a UV photometer as a calibration standard. The
information described here should be adequate to perform the calibration.
However, for more information refer to the Code of Federal Regulations
(Title 40, Part 50, Appendix D) and the EPA's “Technical Assistance
Document for the Calibration of Ambient Ozone Monitors.”
This chapter includes the following sections:
•
•
•
“Calibration Photometer System Preparation”
•
•
“Periodic Zero and Span Checks”
•
“Internal Ozonator Adjustment (Option)”
Equipment Required
The following equipment is required to calibrate the analyzer:
•
Zero air generator
•
Calibration photometer system
Zero Air Generator
Zero air can be obtained either from compressed cylinders or from scrubbed
ambient air. If cylinder air is used, it should be actual and not synthetic. If
ambient air is used, the following compounds must be removed: ozone,
nitric oxide, nitrogen dioxide, sulfur dioxide, and hydrocarbons. The
following scheme is recommended by the EPA in its technical assistance
document:
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...