Servicing
Ozonator Lamp Replacement (Optional)
7-38
Model 49
i
Instruction Manual
Thermo Fisher Scientific
4. At the Flow A Sensor Cal screen, press
to select
Zero
.
The Calibrate Flow A Zero screen appears.
5. Wait at least 10 seconds for the zero reading to stabilize, then press
to save the zero flow value.
6. Reconnect the pump cable to the AC PUMP connector on the
measurement interface board.
7. Connect a calibrated flow sensor at the SAMPLE bulkhead on the rear
panel.
8. Press
to return to the Flow A Sensor Cal screen.
9. At the Flow A Sensor Cal screen, press
to select
Span
.
The Calibrate Flow A Span screen appears.
10. Wait at least 10 seconds for the reading to stabilize, use
and
to enter the flow sensor reading, and press
to
save the value.
11. Install the cover.
Ozonator Lamp
Replacement
(Optional)
Use the following procedure to replace the ozonator lamp.
Equipment Required:
Ozonator lamp
Allen wrench, 7/64-inch
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...