C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-39
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The example that follows shows that the instrument is in
the remote mode.
Send:
mode
Receive:
mode remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The example
that follows sets the instrument to the local mode.
Send:
set mode local
Receive:
set mode local ok
program no
This command reports the analyzer’s model information and program
version number, which will be dependant on the current version.
Send:
program no
Receive:
program no iSeries 49i 01.00.01.074
set layout ack
This command disables the stale layout/layout change indicator (‘*’) that is
attached to each response if the layout has changed.
Send:
set layout ack
Receive:
set layout ack ok
I/O Configuration
analog iout range
channel
This command reports the analog current output range setting for
channels
,
where
channel
must be between 1 and 6, inclusive. The example that follows
reports current output channel 4 to the 4-20 mA range, according to
. This command responds with “feature not enabled” if the I/O
expansion board is not detected.
Send:
analog iout range 4
Receive:
analog iout range 4 2
set analog iout range
channel range
This command sets analog current output
channel
to the
channel range
where
channel
is between 1 and 6 inclusive, and
range
is set according to
. The example that follows sets current output channel 4 to the
0-20 mA range. This command responds with “feature not enabled” if the
I/O expansion board is not detected.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...