Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-95
Analog Input Calibrate Zero
The Analog Input Calibrate Zero screen allows the user to calibrate the zero
state of the selected analog input.
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel >
Calibrate Zero
. (Hook up a voltage source of 0 V to the analog
input channel.)
•
Press
to save the value.
Analog Input Calibrate Full-Scale
The Analog Input Calibration Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
ª
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL: ZERO
DISCONNECT SELECTED INPUT!
SELECTED INPUT: INPUT 1
CURRENTLY: 6.24 V ?
CALIBRATE INPUT TO ZERO
RANGE AVG DIAGS ALARM