Operation
Instrument Controls Menu
3-56
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Minimum and Maximum Value
The MinimumValue screen is used to edit the zero (0) to full-scale (100)
value in percentages for the selected analog output channel. See Table 3-6
for a list of choices. The minimum and maximum output value screens
function the same way. The example below shows the set minimum value
screen.
•
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > Select Channel >
Set Minimum
or
Maximum
Value
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to validate and save the new minimum value.
Table 3-6.
Analog Output Zero to Full-Scale Table
Output
Zero % Value
Full-Scale 100% Value
O
3
Zero (0)
Range Setting
LO O
3
Zero (0)
Range Setting
HI O
3
Zero (0)
Range Setting
Range Status
Recommend not to change the setting for this output
Intensity A
User-set alarm min value
User-set alarm max value
Intensity B
User-set alarm min value
User-set alarm max value
Noise A
Zero (0)
Range Setting
Noise B
Zero (0)
Range Setting
Flow A
User-set alarm min value
User-set alarm max value
Flow B
User-set alarm min value
User-set alarm max value
Pressure
User-set alarm min value
User-set alarm max value
Bench Temp
User-set alarm min value
User-set alarm max value
Lamp Temp
User-set alarm min value
User-set alarm max value
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT: V ALL
CURRENTLY: N/A %
SET TO: 0000.
5
% ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...