Servicing
Optical Bench Temperature Calibration
7-26
Model 49
i
Instruction Manual
Thermo Fisher Scientific
5. Perform the previous steps in reverse to install the new optical bench.
6. Calibrate the instrument. Refer to the “Calibration” chapter in this
manual.
Optical Bench
Temperature
Calibration
Use the following procedure to calibrate the optical bench temperature.
Equipment Required:
Calibrated thermometer or 10K ±1% Resistor
1. Remove the instrument cover.
2. Tape the thermistor (plugged into the measurement interface board) to a
calibrated thermometer.
Note
Since the thermistors are interchangeable to an accuracy of ±0.2 °C,
and have a value of 10K ohms at 25 °C, an alternate procedure is to connect
an accurately known 10K resistor to the thermistor input (TEMP) on the
measurement interface board, and enter the temperature reading.
A 1 °C change corresponds to a ±5% change in resistance, thus this
alternative procedure can be quite accurate as a check; however, it clearly is
not NIST traceable.
3. From the Main Menu, press
to scroll to Service > press
>
to scroll to
Temperature Calibration
> and press
.
WARNING
The service procedures in this manual are restricted to qualified
service representatives.
If the equipment is operated in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...