Servicing
Detector Replacement
7-30
Model 49
i
Instruction Manual
Thermo Fisher Scientific
5. To install the new photometer board, follow previous steps in reverse.
Detector
Replacement
Use the following procedure to replace a detector (
).
Equipment Required:
Detector assembly
Allen wrench, 9/64-inch and 1/16-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the detector electrical connections (CHA and CHB) from
the measurement interface board.
3. Loosen the two screws securing the detector clamping block to the
optical bench and remove the clamping block and detectors
(
Figure 7-12.
Replacing the Detectors
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Clamping Block
Clamping Block Screws (2)
Detectors
Detector Screws (2)