Operation
Alarms Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-73
O
3
Lamp Temperature
The O
3
Lamp Temperature screen (if the ozonator option is installed)
displays the current ozonator lamp temperature and sets the minimum and
maximum alarm limits. Acceptable alarm limits range from 65 to 75 °C. If
the O
3
lamp temperature reading goes beyond either the minimum or
maximum alarm limit, an alarm is activated. The word “ALARM” appears
in the Run screen and in the Main Menu.
•
In the Main Menu, choose Alarms >
O
3
Lamp Temp
.
ALARMS:
ALARMS DETECTED 0
>O3 LAMP TEMP OK
LAMP TEMP OK
BENCH TEMP OK
PRESSURE OK
FLOW A OK
FLOW B OK
ª
RANGE AVG DIAGS ALARM
INTENSITY A OK
INTENSITY B OK
ZERO CHECK OK
SPAN CHECK OK
ZERO AUTOCAL OK
SPAN AUTOCAL OK
OZ LEVEL 1 CHECK OK
OZ LEVEL 2 CHECK OK
OZ LEVEL 3 CHECK OK
OZ LEVEL 4 CHECK OK
OZ LEVEL 5 CHECK OK
O3 CONCENTRATION OK
MOTHERBOARD STATUS OK
INTERFACE STATUS OK
I/O EXP STATUS OK
O3 LAMP TEMPERATURE:
ACTUAL 68.8
o
C
>MIN 60.0
o
C
MAX 80.0
o
C
RANGE AVG DIAGS ALARM
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...