Operation
Alarms Menu
3-78
Model 49
i
Instruction Manual
Thermo Fisher Scientific
Intensity A and B
The Intensity A screen displays the current lamp intensity reading in Cell A
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 45,000 to 150,000 Hertz. If the intensity A reading goes beyond
either the minimum or maximum limit, an alarm is activated. The word
“ALARM” appears in the Run screen and in the Main Menu.
•
In the Main Menu, choose Alarms >
Intensity A
or
Intensity B
.
Min and Max Intensity Limits
The Minimum Intensity A alarm limit screen is used to change the
minimum intensity A alarm limit. The minimum and maximum intensity A
and intensity B screens function the same way.
•
In the Main Menu, choose Alarms > Select Intensity >
Min
or
Max
.
•
Use
and
to increment or decrement the numeric value.
•
Press
to save set to value as actual value.
FLOW A:
ACTUAL MIN: 0.400 LPM
SET MIN TO: 0.500 LPM ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
INTENSITY A:
ACTUAL 98425 Hz
>MIN 45000 Hz
MAX 150000 Hz
RANGE AVG DIAGS ALARM
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...