
Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Instruction Manual
3-91
•
In the Main Menu, choose Service > Select Flow Calibration A or B >
Zero
.
•
Press
to save the current flow reading as the zero reading.
Calibrate Flow Span
The Calibrate Flow A Span screen allows the user to view and set the flow
sensor calibrate span point. The calibrate Flow B Span screen functions the
same way.
Note
An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration.
•
In the Main Menu, choose Service > Select Flow Calibration A or B >
Span
.
•
Use
,
,
and
to move and change the value.
•
Press
to save set to value as actual value.
Restore Default Flow Calibration
The Restore Default Flow Calibration A screen allows the user to reset the
flow calibration configuration values to factory defaults. The Restore
Default Calibration B screen functions the same way.
CALIBRATE FLOW A ZERO:
CURRENTLY: 0.608 LPM
SET TO: 0.000 LPM ?
DISCONNECT PUMP AND
SAVE ZERO PRESSURE
RANGE AVG DIAGS ALARM
CALIBRATE FLOW A SPAN:
CURRENTLY: 0.608 LPM
SET TO: 0.80
0
LPM ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...