Calibration
Calibration Procedure
Thermo Fisher Scientific
Model 49
i
Instruction Manual
4-9
•
after performing a major component disassembly
Connect Instrument
Connect the Model 49
i
to the manifold on the output of the ozonator, as
shown in
. If an optional sample line filter is used, the calibration
must be performed through this filter. Ensure that the flow rate into the
output manifold is greater than the total flow required by the calibration
photometer, analyzer, and any other flow demand connected to the
manifold.
Figure 4-2.
Model 49
i
Connected to Calibration Photometer and External
Ozonator
Zero Adjust
Use the following procedure to adjust zero.
1. Allow sufficient time for the instrument and the calibration photometer
to warm up and stabilize.
2. With the zero air supply ON, but the ozonator OFF, allow the
instrument to sample zero air until a stable response is obtained.
3. From the Main Menu choose
Calibration
.
4. From the Calibration menu choose
Calibrate Zero
. Press
to
perform a zero calibration. Press
to return to the Run screen.
Model 49
i
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...