C-Link Protocol Commands
Record Layout Definition
Thermo Fisher Scientific
Model 49
i
Instruction Manual
B-47
N—Indicates a button which only sends the subsequent command to
the instrument. No user-prompting happens.
The following string through an optional ‘|’ or the end of the line is the
command which is to be sent to the instrument upon the completion of the
button selection. The command string should normally contain print-style
formatting to include the user input. If a ‘|’ is present, it indicates a
command which is sent to the instrument upon successful completion of
the button command to update the value field.
This is not currently used.
Examples
Some examples ('\n' is the C syntax for an end-of-line character):
'
Concentrations\n
'
This is a single text-only line.
'
\n
'
This is a single blank line.
'
O3:3s\n
'
This is a line which appears slightly indented. The text field is '
O3
', the value
is taken from the third element of the data response, and interpreted as a
string.
'
O3:18sBd.ddd;set no coef %s\n
'
This is a line which also appears slightly indented. The next field is also '
O3
',
but the value is taken from the eighteenth element of the data response,
again interpreted as a string. A button appears on this line which, when
pressed, pops up an input dialog which will state "Please enter a new value
for
O3
using a d.ddd format." The string entered by the user is used to
construct the output command. If the user enters, for example, '1.234', the
constructed command will be 'set no coef 1.234'.
'
O3:21f{Code_0 Code_1 Code_2 Code_3 Code_4 Code_5 Code_6 Code_7
Code_8 Code_9 Code_10 Code_11}Lset range no %d\n
'
This is a line which appears slightly indented, the title is again '
O3
', and the
value the twenty-first element of the data response, interpreted as a
floating-point number. There is a no-translation button which creates a
selection list of twelve "Code nn" options. The number of the user selection
is used to create the output command.
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...