Thermo Fisher Scientific
Model 49
i
Instruction Manual
D-1
Appendix D
Geysitech (Bayern-Hessen)
Protocol
This appendix provides a description of the Geysitech (Bayern-Hessen or
BH) Protocol Interface and is supported both over RS-232/485 as well as
TCP/IP over Ethernet.
The Geysitech Commands that are implemented are explained in detail in
this document. The Geysitech protocol support for the
i
Series enables the
user to perform the functions of reading the various concentrations and to
trigger the instrument to be in sample/zero/span mode if valid for that
instrument. This is achieved by using the supported Geysitech commands
listed below.
For details of the Model 49
i
Geysitech Protocol specification, see the
following topics:
“Serial Communication Parameters”
“TCP Communication Parameters”
Serial Communication
Parameters
The following are the communication parameters that are used to configure
the serial port of the
i
Series to support Geysitech protocol.
Number of Data bits : 8
Number of Stop bits : 1
Parity :
None
Data rate
: from 1200-115200 Baud (9600 is default)
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...