Servicing
Optical Bench Replacement
Thermo Fisher Scientific
Model 49
i
Instruction Manual
7-25
Screwdriver, flatblade
Use the following procedure to replace the optical bench.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the plumbing and electrical connections.
3. Remove the two screws from the photometer lamp access hole cover on
the rear panel and remove the cover.
4. Loosen the captive screws (4) securing the optical bench to the floor
) and remove the bench by first moving the
photometer slightly towards the rear through the lamp access hole and
then lifting up and towards the front.
Figure 7-11.
Replacing the Optical Bench
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Photometer Lamp
Lamp Heater
Access Port Cell B IN
Absorption Cell B
Access Port Cell B OUT
Detector B
Access Port Cell A IN
Input Block
Absorption Cell A
Access Port Cell A OUT
Output Block
Detector A
Captive Screws (4)
Summary of Contents for 49i
Page 1: ...Model 49i Instruction Manual UV Photometric O3 Analyzer Part number 102434 00 10Jan2008 ...
Page 4: ......
Page 6: ......
Page 14: ...Contents xiv Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Figures xvi Model 49i Instruction Manual Thermo Fisher Scientific Flag Status B 11 ...
Page 36: ...Installation Startup 2 10 Model 49i Instruction Manual Thermo Fisher Scientific ...
Page 184: ...Troubleshooting Service Locations 6 22 Model 49i Instruction Manual Thermo Fisher Scientific ...