Servicing
Analog Output Testing
7-12
Model 49
i
Instruction Manual
Thermo Fisher Scientific
5. Install a new fan following the previous steps in reverse.
Figure 7-5.
Replacing a Fan
Analog Output Testing
The analog outputs should be tested if the concentration value on the front
panel display disagrees with the analog outputs. To check the analog
outputs, you connect a meter to an analog output channel (voltage or
current) and compare the meter reading with the output value displayed on
the Test Analog Outputs screen.
Equipment Required:
Multimeter
Use the following procedure to test the analog outputs.
1. Connect a meter to the channel to be tested.
shows the
identifies the associated channels.
2. From the Main Menu, press
to scroll to Diagnostics, press
>
to scroll to Test Analog Outputs, and press
.
The Test Analog Outputs screen displays.
3. Press
to scroll to the channel corresponding to the rear panel
terminal pins where the meter is connected, and press
.
The Set Analog Outputs screen displays.