Servicing
Analog Output Testing
Thermo Fisher Scientific
Model 49
i
Instruction Manual
7-13
4. Press
to set the output to zero.
The Output Set To line displays Zero.
5. Check that the meter is displaying a zero value. If the meter reading
differs by more than one percent, the analog outputs should be
adjusted. Refer to the “Analog Output Calibration” procedure that
follows.
6. Press
to set the output to full scale.
The Output Set To line displays Full Scale.
7. Check that the meter is displaying the full scale value. If the meter
reading differs by more than one percent, the analog outputs should be
adjusted. Refer to the “Analog Output Calibration” procedure that
follows.
8. Press
to reset the analog outputs to normal.