Teledyne API Model T360/T360M Operation Manual
Getting Started
63
3.8.1.2. O
2
CALIBRATION METHOD
STEP 1 – SET O
2
SPAN GAS CONCENTRATION :
Set the expected O
2
span gas concentration.
This should be equal to the percent concentration of the O
2
span gas of the
selected reporting range (default factory setting = 20.8%; the approximate O
2
content of ambient air).
SAMPLE
GAS TO CAL:CO2
CO2
O2
ENTR EXIT
SAMPLE RANGE = 500.000 PPM CO2 =XXX.X
< TST TST > CAL
SETUP
M-P CAL A1:NXCNC1 =100PPM NOX=X.XXX
<TST TST> ZERO SPAN
CONC
EXIT
SAMPLE
GAS TO CAL:O2
NOX
O2
ENTR
EXIT
M-P CAL O2 SPAN CONC:20.8%
0
2
0
.8
0
ENTR EXIT
The O
2
span concentration value automatically defaults to
20.8 %
.
If this is not the the concentration of the span gas being
used, toggle these buttons to set the correct concentration
of the O
2
calibration gases.
EXIT
ignores the new
setting and returns to
the previous display.
ENTR
accepts the new
setting and returns to
the previous menu.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...