Teledyne API Model T360/T360M Operation Manual
Calibration Procedures
183
5.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected
SETUP C.4
RANGE TO CAL: HIGH
<SET
EDIT
EXIT
EXIT
returns to the
PRIMARY SETUP
Menu
SETUP C.4
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
SETUP C.4
RANGE TO CAL: LOW
LOW
HIGH
ENTR
SETUP
SETUP C.4
RANGE TO CAL: LOW
<SET
EDIT
EXIT
NOTE
In order to automatically calibrate both the HIGH and LOW ranges, you must set up a separate
sequence for each.