Operating Instructions
Teledyne API Model T360/T360M Operation Manual
104
Example: If all passwords are enabled, the following menu button sequence
would be required to enter the SETUP menu:
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS: 0
0 0 0
ENTR EXIT
prompts for password
number
See Table 6-8 for
Passwords and Levels
Example: this
password enables the
SETUP mode
SAMPLE
ENTER SETUP PASS: 0
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
Press individual
buttons to set
numbers
Note that the instrument still prompts for a password when entering the VARS
and DIAG menus, even if passwords are disabled, but it displays the default
password (818) upon entering these menus. The user only has to press ENTR to
access the password-protected menus but does not have to enter the required
number code.