Calibration Procedures
Teledyne API Model T360/T360M Operation Manual
182
To program the Sequence:
Toggle
buttons to
set
number of
days
between
procedures
(1-367)
Toggle
buttons to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP X.X
DELTA DAYS: 1
<SET SET>
EDIT
EXIT
SETUP X.X
DELTA DAYS: 1
0 0 2
ENTR
EXIT
SETUP X.X
DELTA DAYS:2
<SET
SET>
EDIT
EXIT
SETUP X.X
DELTA TIME00:00
<SET SET>
EDIT
EXIT
SETUP X.X
DELTA TIME: 00:00
0 0 :3 0
ENTR
EXIT
SETUP X.X
DELTA TIEM:00:30
<SET
SET>
EDIT
EXIT
SETUP X.X
STARTING TIME:14:15
<SET
SET>
EDIT
EXIT
EXIT
returns
to the
SETUP
Menu
SETUP X.X
DURATION:15.0 MINUTES
<SET SET>
EDIT
EXIT
SETUP X.X
DURATION 15.0MINUTES
3 0 .0
ENTR
EXIT
SETUP X.X
DURATION:30.0 MINUTES
<SET
SET>
EDIT
EXIT
Toggle
buttons to
set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP X.X
CALIBRATE: OFF
<SET SET>
EDIT
EXIT
SETUP X.X CALIBRATE: OFF
ON
ENTR
EXIT
SETUP X.X
CALIBRATE: ON
<SET SET> EDIT
EXIT
Toggle
button
between
Off and
ON
SETUP X.X
SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT
MODE
SET
EXIT
Sequence
Delta Time
MODE Delta Days
SETUP X.X
PRIMARY SETUP MENU
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SEQ 1) DISABLED
NEXT
MODE
EXIT
SETUP X.X
SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP X.X
MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO
PREV
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO–SPAN
PREV
NEXT
ENTR
EXIT
SETUP X.X
SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT
MODE
SET
EXIT
SETUP X.X
TIMER ENABLE: ON
SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
0
4 SEP
0
3
ENTR
EXIT
Toggle
buttons to set
day, month &
year:
Format :
DD-MON-Y
Y
SETUP X.X
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT
EXIT
Default
value is
ON
SETUP X.X
STARTING DATE: 04–SEP–03
<SET
SET>
EDIT
EXIT
SETUP X.X
STARTING TIME:00:00
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING TIME:00:00
1 4 : 1 5
ENTR
EXIT/
Toggle buttons
to set time:
Format : HH:MM
This is a 24 hr
clock .
PM hours are
13 – 24.
Example
2:15 PM = 14:15
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL CALZ CALS
SETUP
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...