Teledyne API Model T360/T360M Operation Manual
Operating Instructions
101
4.8.6. Range Units
The T360 can display concentrations in parts per billion (10
9
mols per mol,
PPB
),
parts per million (10
6
mols per mol,
PPM
), micrograms per cubic meter (µg/m
3
,
UG
), milligrams per cubic meter (mg/m
3
,
MG
) or percent (volume CO
2
/volume
sample gas,
%
). Changing units affects all of the display, analog outputs, COM
port and DAS values for all reporting ranges regardless of the analyzer’s range
mode.
NOTE
Concentrations displayed in mg/m3 and ug/m3 use 0
C, 760 mmHg for Standard Temperature and
Pressure (STP). Consult your local regulations for the STP used by your agency.
Conversion factors from volumetric to mass units are:
CO
2
: ppb x 1.96 = µg/m3; ppm x 1.96 = mg/m3
To change the concentration units:
SETUP X.X
CONC UNITS: %
PPM PPB
UGM
MGM %
ENTER EXIT
SAMPLE RANGE = 500.00 PPB CO2=X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
CONC UNITS: PPM
PPM PPB
UGM
MGM
%
ENTER
EXIT
Select the preferred
concentration unit.
EXIT
returns
to the main menu.
ENTR
accepts
the new unit,
EXIT
returns
to the
SETUP
menu.
NOTE
Once the units of measurement have been changed the unit MUST be recalibrated, as the
“expected span values” previously in effect will no longer be valid. Simply entering new expected
span values without running the entire calibration routine is not sufficient.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...