4.13.4.5. AIN CALIBRATION
This is the sub-menu to conduct the analog input calibration. This calibration
should only be necessary after major repair such as a replacement of CPU,
motherboard or power supplies. Activate the ANALOG I/O
CONFIGURATION MENU (see Section 4.13.1), then press:
Exit at any time to
return to the main
DIAG
menu
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
Continue pressing
SET?
until …
DIAG AIO
AIN CALIBRATED:
NO
< SET
SET>
CAL
EXIT
DIAG AIO
AIN CALIBRATED:
YES
< SET
SET> CAL
EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
STARTING FROM ANALOG I / O CONFIGURATION MENU
DIAG AIO
CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
4.13.4.6. CONFIGURING ANALOG INPUTS (OPTION) CHANNELS
To configure the analyzer’s external analog inputs option, define for each
channel
:
gain (number of units represented by 1 volt)
offset
(volts)
engineering units to be represented in volts (each press of the touchscreen
button scrolls the list of alphanumeric characters from A-Z and 0-9)
whether to display the channel in the Test functions
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...