4.8.4. Dual Range Mode (DUAL)
Selecting Dual Range mode allows the
A1
and
A2
outputs to be configured with
different reporting ranges. The analyzer software calls these two ranges low and
high. The low range setting corresponds with the analog output labeled
A1
on the
Rear Panel of the instrument. The high Range Setting corresponds with the
A2
output. While the software names these two ranges low and high, they do not
have to be configured that way. For example: The low range can be set for a span
of 0-1000 ppm while the high range is set for 0-500 ppm.
In
DUAL
range mode the
RANGE
test function displayed on the front panel will
be replaced by two separate functions:
RANGE1
: The range setting for the
A1
output.
RANGE2
: The range setting for the
A2
output.
To set the ranges press following menu button sequence
.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL
DUAL
AUTO
ENTR EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE: 500.0 Conc
0 0 5
0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE
:
500.0 Conc
0 0
1
0 0 .0
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
EXIT
Returns
to the Main
SAMPLE
Display
SETUP X.X
RANGE MODE:
DUAL
SNGL
DUAL AUTO
ENTR
EXIT
Toggle the
numeral buttons
to set the upper
limit of each
range.
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
When the instrument’s range mode is set to
DUAL
the concentration field in the
upper right hand corner of the display alternates between displaying the low
range value and the high range value. The concentration currently being
displayed is identified as follows: C1 = Low (or
A1
) and C2 = High (or
A2
).
NOTE
In DUAL range mode the LOW and HIGH ranges have separate slopes and offsets for computing
CO
2
concentration.
The two ranges must be independently calibrated.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...