Calibration Procedures
Teledyne API Model T360/T360M Operation Manual
172
STEP THREE: Perform the Zero/Span Calibration Procedure:
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
ENTR
CONC
EXIT
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
CAL
SETUP
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =XXX.X
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL RANGE = 500.000 PPM CO2 =XXX.X
< TST TST > ENTR CONC
EXIT
Wait until
STABIL
falls below
1.0 ppm.
This may take several
minutes.
The
SPAN
button
appears.
If either the
ZERO
or
SPAN
buttons fail to
appear see Section 11
for troubleshooting tips.
SAMPLE* RANGE = 500.000 PPM CO2 =XXX.X
< TST
TST
>
CAL
SETUP
SAMPLE
STABIL=XXX.X PPM
CO2 =XXX.X
< TST TST >
CAL
SETUP
Set the Display to show the
STABIL
test function.
This function calculates the
stability of the CO
2
measurement
Press
ENTR
to changes the
OFFSET
&
SLOPE
values for the
CO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
STABIL
may jump
significantly.
Wait until it falls back
below
1.0 ppm
This may take several
minutes.
Press
ENTR to
change the
OFFSET
&
SLOPE
values for the
CO
2
measurements.
Press
EXIT
to leave the calibration
unchanged and return to the
previous menu.
If the
ZERO
or
SPAN
buttons are not displayed, this means that the
measurement made during that part of the procedure is too far out of the
allowable range to do allow a reliable calibration. The reason for this must be
determined before the analyzer can be calibrated. See Section 8 for
troubleshooting tips.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...