Teledyne API Model T360/T360M Operation Manual
Operating Instructions
139
To automatically calibrate a single analog channel, activate the ANALOG I/O
CONFIGURATION MENU (see Section 4.13.1), then press:
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
<SET
CAL
EXIT
DIAG AIO
AUTO CALIBRATING CONC_OUT_2
DIAG AIO
CONC_OUT_2 CALIBRATED:
YES
<SET CAL
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
<
SET>
CAL
EXIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 AUTO CAL: ON
< SET
SET>
EDIT
EXIT
EXIT
to Return
to the main
Sample Display
Press
SET>
to select the
Analog Output channel to
be configured. Then Press
EDIT
to continue
07272B DCN6552