Teledyne API - T360/T360M, 360E/360EM Appendix A Menu Trees (05233C DCN6552)
A-24
APPENDIX A-3:
300 Series
Warnings and Test Measurements, Revision L.8
Name
1
Message Text
Description
Warnings
WSYSRES SYSTEM
RESET
Instrument was power-cycled or the CPU
was reset.
WDATAINIT
DATA INITIALIZED
Data storage was erased.
WCONFIGINIT CONFIG
INITIALIZED Configuration storage was reset to factory
configuration or erased.
WCONCALARM1
CONC ALARM 1 WARN
Concentration limit 1 exceeded.
WCONCALARM2
CONC ALARM 2 WARN
Concentration limit 2 exceeded.
WSOURCE SOURCE
WARNING Reference reading minus dark offset
outside of warning limits specified by
REF_SOURCE_LIMIT
variable.
WAUTOZERO
4, 5
AZERO WARN 1.001
Auto-reference ratio below limit specified
by
ZERO_LIMIT
variable.
WBENCHTEMP
BENCH TEMP WARNING
Bench temperature outside of warning
limits specified by
BENCH_SET
variable.
WWHEELTEMP
WHEEL TEMP WARNING
Wheel temperature outside of warning
limits specified by
WHEEL_SET
variable.
WO2CELLTEMP
10
O2 CELL TEMP WARN
O
2
sensor cell temperature outside of
warning limits specified by
O2_CELL_SET
variable.
WSAMPFLOW
6
SAMPLE FLOW WARN
Sample flow outside of warning limits
specified by
SAMP_FLOW_SET
variable.
WSAMPPRESS
SAMPLE PRESS WARN
Sample pressure outside of warning limits
specified by
SAMP_PRESS_SET
variable.
WSAMPTEMP
13
SAMPLE TEMP WARN
Sample temperature outside of warning
limits specified by
SAMP_TEMP_SET
variable.
WPURGEPRESS
9
PURGE PRESS WARN
Purge pressure outside of warning limits
specified by
PURGE_PRESS_SET
variable.
WBOXTEMP BOX
TEMP
WARNING Internal box temperature outside of
warning limits specified by
BOX_SET
variable.
WBOXTEMP2
4
BOX
TEMP2
WARNING Internal box temperature #2 outside of
warning limits specified by
BOX2_SET
variable.
WOVENTEMP
11
OVEN TEMP WARNING
Oven temperature outside of warning
limits specified by
OVEN_SET
variable.
WPHOTOTEMP PHOTO
TEMP
WARNING Photometer temperature outside of
warning limits specified by
PHOTO_TEMP_SET
variable.
WDYNZERO
CANNOT DYN ZERO
Contact closure zero calibration failed
while
DYN_ZERO
was set to
ON
.
WDYNSPAN
CANNOT DYN SPAN
Contact closure span calibration failed
while
DYN_SPAN
was set to
ON
.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...