
Operating Instructions
Teledyne API Model T360/T360M Operation Manual
102
4.8.7. Dilution Ratio
The dilution ratio is a software option that allows the user to compensate for any
dilution of the sample gas before it enters the sample inlet. Using the dilution
ratio option is a 4-step process:
1. Select reporting range units (Section 4.8.6).
2. Select the range: Use the procedures in Sections 4.8.2 – 4.8.5. Make sure
that the
SPAN
value entered is the maximum expected concentration of the
undiluted calibration gas and that the span gas is either supplied through the
same dilution inlet system as the sample gas or has an appropriately lower
actual concentration. For example, with a dilution set to 100, a 10 ppm gas
can be used to calibrate a 1000 ppm sample gas if the span gas is not routed
through the dilution system. On the other hand, if a 1000 ppm span gas is
used, it needs to pass through the same dilution steps as the sample gas.
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluting
gas and 1 part of sample gas):
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP C.3
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
DIL
EXIT
DIL
only appears
if the dilution ratio
option has been
installed
SETUP C.3
DIL FACTOR: 1.0 GAIN
0 0 0 1 .0
ENTR
EXIT
Toggle these buttons to set the
dilution factor.
This is the number by which the
analyzer will multiply the CO
2
concentrations of the gas passing
through the reaction cell.
SETUP C.3
DIL FACTOR: 20.0 GAIN
0 0 2
0 .0
ENTR
EXIT
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
The analyzer multiplies the measured gas concentrations with this dilution factor
and displays the result.
NOTE
Once the above settings have been entered, the instrument needs to be recalibrated using one of
the methods discussed in Section 5.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...