Teledyne API Model T360/T360M Operation Manual
Calibration Procedures
173
5.3. MANUAL CALIBRATION CHECKS
Informal calibration checks, which only evaluate but do not alter the analyzer’s
response curve, are recommended as a regular maintenance item and in order to
monitor the analyzer’s performance. To carry out a calibration check rather than
a full calibration, follow these steps.
STEP ONE
: Connect the sources of zero air and span gas as shown in Figures
7.1 or 7.2.
STEP TWO
: Perform the zero/span calibration check procedure:
SAMPLE
STABIL=XXX.X PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
ACTION:
Record the CO
2
concentration
reading.
ACTION:
Supply span gas to the instrument
SAMPLE
STABIL=XXX.X PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
ACTION:
Record the CO
2
concentration
reading.
Wait until
STABIL
is
below
1.0 ppm.
This may take
several minutes.
The value of
STABIL
may jump
significantly.
Wait until it falls
below
1.0 ppm. This
may take several
minutes.
SAMPLE
STABIL=XXX.X PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
The
SPAN
key appears during the transition from zero to
span. You may see both
SPAN
and
ZERO
buttons.
SAMPLE RANGE = 500.0 PPM
CO2=X.XXX
< TST TST >
CAL
SETUP
Scroll the display to the
STABIL
test function.
ACTION:
Supply the instrument with zero gas.
5.4. MANUAL CALIBRATION WITH ZERO/SPAN VALVES
There are four different zero/span valve option configurations (see Figure 5-3
through Figure 5-6). They all operate identically, differing only in the method
used to supply calibration gas to the Analyzer.
STEP ONE:
Connect the sources of Zero Air and Span Gas as shown below.
Figure 5-3 through Figure 5-6 show the proper pneumatic connections for T360s
with various optional internal valve sets installed.
07272B DCN6552
Summary of Contents for T360
Page 2: ......
Page 188: ...This page intentionally left blank 07272B DCN6552 ...
Page 264: ...07272B DCN6552 ...
Page 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Page 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Page 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Page 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Page 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Page 333: ...07272B DCN6552 D 7 ...
Page 338: ...07272B DCN6552 D 12 ...