[NTSC] : Q
[PAL] : U
3-3-12. Q (U) Level Adjustment
Equipment:
Vectorscope
Test point:
TEST OUT connector/side panel
Preparations
.
Extend the EN-141 board with the extension board (EX-762 board assembly or EX-464.)
.
Connect the vectorscope to the TEST OUT connector/side panel.
.
Disconnect the triax cable or optical fiber cable from CCU.
.
Supply power from the AC adaptor (AC-550/550CE).
.
OUTPUT switch/side panel
→
BARS
.
S600-1 (Q/U ON/OFF) /EN-141 board
→
OFF
.
Turn PHASE control on the vectorscope so that the beam line of Q (U) signal matches the Q (U) axis.
Adjustment procedure
1.
Adj. point:
1
RV600 (Q/U LEV) /EN-141 board
Specification:
The beam spot of Q (U) signal fit the scale
Setting after adjustment
S600-1 (Q/U ON/OFF) /EN-141 board
→
ON
3-3. Encoder System Adjustment