5 Operation
Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
94
64 08 137 D3610
D3610.201.02.09.02 11.2017
5.8
Releasing the exposure
ü
The exposure parameters for the X-ray tube assembly are set, see
"Select the exposure parameters for the X-ray tube
assembly" [→ 91].
ü
The sensor is positioned in the patient's mouth with the
corresponding auxiliary exposure equipment, see "Position the
sensor" [→ 74].
1.
Ensure that the WiFi interface is ready for exposures. The LED
must be lit up from dimmed to bright ("breathing") on the WiFi
interface, see "Determine unit status" [→ 66].
2.
Ensure that SIDEXIS is ready for exposures. The green display
must be flashing in the Exposure readiness window.
3.
Check that the X-ray tube assembly is in the correct position and
take the X-ray exposure.
4.
Remove the sensor from the hygienic protective sleeve. For this
follow the instructions in section "Remove the hygienic protective
sleeve from the sensor" [→ 95]. The used sensor holder tab and
hygienic protective sleeve must be disposed of after the
examination. The guide rod and localizer ring must be cleaned and
sterilized. The sensor must be wipe disinfected before using it with
another patient.
5.
Following the X-ray exposure, carefully remove the WiFi interface
and the sensor to prevent them from falling down.
6.
Continue with image processing in SIDEXIS.
CAUTION
The WiFi interface may heat to up to 48°C in the charger in normal
operating conditions
With increased temperatures we recommend using patient clips so as
to avoid direct bodily contact with the patient.
If the temperature of the WiFi interface is considerably higher than the
ambient temperature (difference in temperature > 8°C) and is not
pleasant to handle then please inform the service engineer
responsible. In this case, the device should no longer be used for
patient exposures.