6 Maintenance and inspection
Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
106
64 08 137 D3610
D3610.201.02.09.02 11.2017
2.
Screw the sensor cable to the sensor using the screws supplied.
Initially only tighten the screws until a light resistance can be felt.
After this tighten both screws carefully.
3.
Place a silicon protective cap on each screw.
4.
Cover the screw heads with a new protective screw cap. Press the
cap into the plug of the sensor cable until it locks in place.
Ä
The cable replacement is now completed. The sensor can be used
again.
6.4
Replace the WiFi interface storage battery
The storage battery must be replaced once it has been exhausted and
there is only enough charge remaining for a few X-ray exposures.
The replacement set "XIOS XG WiFi Battery" REF 64 04 375 is required
for this. This contains a new storage battery and a suitable screwdriver.
CAUTION
Hazards may be caused by lithium-ion storage batteries.
Improper use may result in fire as well as personal injury and damage
to property.
➢ Never short-circuit the storage battery!
➢ Do not dismantle or convert the battery pack.
➢ Do not expose the storage battery to temperatures exceeding
66°C (150°F). Keep away from fire and heat.
➢ Do not expose the storage battery to heavy impacts or vibrations.
➢ Keep the storage battery dry.
Please also observe the general safety instructions, see section
"Storage battery" [→ 12].