D3610.201.02.09.02 11.2017
6.3
Replace the sensor cable
The sensor cable can be replaced with a new one in the event of
damage.
A "sensor cable" replacement set is required for this, see Consumables
and spare parts [→ 108]"Consumables and spare parts". This contains
the required spare parts and tool:
● Sensor cable
● Screwdriver
● 2 protective screw caps
● 2 Elastomer strips
● 4 flathead screws
NOTICE
The electric plug contacts of the sensor are exposed during
installation.
The sensor may be destroyed by the electrostatic discharge at the
contacts.
➢ Before installation discharge at a conductive grounded item, e.g.,
a water faucet or a bare heating pipe.
NOTICE
No dirt or moisture must be allowed to get into the exposed plug
contacts.
Dirt causes contacts faults, moisture may lead to a short circuit.
➢ Wash and dry your hands. Do not wear powdered gloves. The
powder could become deposited in the plug contacts of the
sensor.
➢ Place the sensor on a clean, dry and stable surface.
Unscrew a damaged cable from the sensor
1.
Disconnect the sensor from the XIOS XG system.
2.
With e.g. a dental instrument remove the protective screw cap from
the rear side of the sensor.