4.12.3
Sensor Upgrade
IMPORTANT
The WiFi interface may not be in the charger tray for the following
steps.
1.
Switch the WiFi interface on and wait 10 seconds until the LED
flashes.
2.
Connect the sensor that is to receive an upgrade to the WiFi
interface.
3.
Click on the "Wireless Monitor" symbol in the info area of the
Windows taskbar (Systray) and select the "WiFi Interface
Management" entry.
Ä
A list opens with the WiFi interfaces found in the wireless
network. It can take up to 25 seconds until all WiFi interfaces
are listed. The WiFi interfaces are named after the unit serial
number in the list.
4.
Select the WiFi interface that is to receive an upgrade from the list.
Ensure that the WiFi interface is
not paired with the PC.
5.
Right-click on the desired WiFi interface and select the entry "Open
Device Web Interface" in the context menu.
Ä
The "Web Server and Upgrade Utility" page is displayed.
6.
Click the "Firmware Upgrade" tab.