Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
8 Electromagnetic compatibility
64 08 137 D3610
D3610.201.02.09.02 11.2017
117
Power rating of the transmitter
[W]
Working clearance according to transmission frequency [m]
150 kHz - 80 MHz
80 MHz - 800 MHz
800 MHz - 2.5 GHz
d= [1.2] √P
d= [1.2] √P
d= [2.3] √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
The recommended safety distance
d
in meters (m) can be determined
for transmitters, whose maximum power rating is not specified in the
above table, using the equation that belongs to the corresponding
column, wherein
P
is the maximum power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
Note 1
The higher frequency range applies at 80 MHz and 800 MHz.
Note 2
These guidelines may not apply in all cases. The propagation of
electromagnetic waves is influenced by their absorption and reflection
by buildings, objects and persons.