4 Installation
Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
54
64 08 137 D3610
D3610.201.02.09.02 11.2017
4.12
Web Server and Upgrade Utility page
Functional scope overview
The Web Server and Upgrade Utility page is split up into multiple tabs
which cover the following functions:
● "Status"
Status display for the WiFi interface
● "Firmware Upgrade"
Firmware upgrade installations
● "Diagnostic"
Diagnostic tests completed
● "Settings"
User-defined settings display
"Status"
The software version and date information, WiFi interface serial number
and sensor information are displayed in "Status" (where a sensor is
connected to the WiFi interface).
"Firmware Upgrade"
Firmware upgrades contain the latest enhancements for your product.
We recommend that you install these urgently. The upgrades are easy
to complete, meaning that the user is able to manage this themselves.
There are three different types of upgrade:
● Flash upgrade for the WiFi interface
● FPGA upgrade for the WiFi interface (Field-programmable gate
array)
● Firmware upgrade of the sensor
Although we recommend that the WiFi interface is plugged into the
charger for Flash and FPGA upgrades, these upgrades can also be
completed with the device outside the charger. In this case the charge
state for the storage battery must be 50% or higher in order to ensure
that the power supply is not interrupted during the upgrade process.
The high amount of power required in the upgrade process can quickly
drain the storage battery, resulting in an incomplete upgrade and
preventing the device from restarting.
"Diagnostic"
The diagnostic tests include a bandwidth test which measures how
quickly an image can be transferred from the WiFi interface to the PC
and displayed.
"Settings"
The Settings include the user-defined settings.