3 System description
Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
28
64 08 137 D3610
D3610.201.02.09.02 11.2017
XIOS XG Select Sensors
Technology:
CMOS-APS (Active Pixel Sensor)
Physical pixel size:
15 µm, image acquisition in 30 µm
Line pairs:
16.7 Lp at 30 µm
Measured resolution:
16 Lp/mm
Theoretical resolution:
16.7 Lp/mm
Active sensor area:
Size 0 - sensor = 18 x 24 mm
Size 1 - sensor = 20 x 30 mm
Size 2 - sensor = 25.6 x 36 mm
External dimensions:
Size 0 - sensor = 23.5 x 32 x 6.3 mm
Size 1 - sensor = 25.3 x 38.4 x 6.3 mm
Size 2 - sensor = 31.2 x 43.9 x 6.3 mm
Cable length:
max. 2.70 m
XIOS XG Supreme Sensors
Technology:
CMOS-APS (Active Pixel Sensor)
Physical pixel size:
15 µm, image acquisition in 15 µm
Line pairs:
33.3 Lp at 15 µm
Measured resolution:
28 Lp/mm
Theoretical resolution:
33.3 Lp/mm
Active sensor area:
Size 0 - sensor = 18 x 24 mm
Size 1 - sensor = 20 x 30 mm
Size 2 - sensor = 25.6 x 36 mm
External dimensions:
Size 0 - sensor = 23.6 x 32 x 7.5 mm
Size 1 - sensor = 25.4 x 38.3 x 7.5 mm
Size 2 - sensor = 31.2 x 43 x 7.5 mm
Cable length:
max. 2.70 m
Power supply unit
Input:
100-240 V ~50/60 Hz
500 mA
Output:
6.5 V
1.5 A
Operating and transport conditions
Mode of operation:
Continuous operation
Additional information:
The unit must not be used near flammable
anesthesia gas mixtures containing air, oxy-
gen or nitrogen oxide.
Transport and storage
conditions:
Temperature: -40 °C (-40 °F) to 70 °C (158
°F)
Relative humidity: 20 % to 85 %
Air pressure: 500 to 1060 hPa