Sirona Dental Systems GmbH
Gebrauchsanweisung und Installation
5 Operation
64 08 137 D3610
D3610.201.02.09.02 11.2017
95
5.9
Removing the hygienic protective sleeve from
the sensor
5.9.1
With single-use sensor holder system
Have the sensor holder tab and the sensor holder glued onto the
hygienic protective sleeve. This makes it easier to remove the hygienic
protective sleeve.
NOTICE
The sensor cable is sensitive to mechanical influences.
The cable may become damaged or may wear out prematurely.
➢ Do not pull on the sensor cable when pulling the sensor out of the
hygienic protective sleeve. Slide the sensor further out of the
hygienic protective sleeve with your thumb as described below.
➢ Avoid bending, creasing or rotating the cable or exposing it to
other strains.
1.
Grasp the guide rod in one hand so that you can touch the side of
the sensor facing away from the sensor cable with your thumb.
2.
Carefully push the sensor out of the part of the hygienic protective
sleeve that is glued to the sensor holder tab with your thumb.
Summary of Contents for XIOS XG Select
Page 119: ......