76
CLINITEK Analyzer Operator’s Guide
Troubl
eshooting
The following table contains the issues that can occur when you
operate the analyzer.
Calling for Assistance
If your CLINITEK analyzer displays corrective actions for a
detected problem, carry out the instructions provided before you call
for assistance. If your actions do not correct the problem or the
instructions do not display, contact your local technical support
provider or distributor.
Technical Support
When you call for assistance about an error message, have the
following items ready. These items help the technical support
representative work on the issue as quickly as possible.
•
Error number
•
Completed problem list (see
Description
Action
Display shows dashes
Dashes in a field indicate where you
disabled a an option.
Dashes also display when you exclude
urinalysis tests for chemistries from the
test results.
You can write the information on the
blank lines in the test result printout, if
needed.
Test table movement is
irregular or slow
Heavy buildup of dried urine on the test
table.
•
Clean the test table and insert as
described in Section 4,
,
the Test Table and Test Table Insert
Low battery power.
•
Replace the batteries as described
in Section 4,
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...