44
CLINITEK Analyzer Operator’s Guide
Op
er
at
ions
•
Sample interference notes (if enabled in Instrument Set Up)
For instructions on how to set up the analyzer so that you can print
the results automatically or send the results to a computer, see
Section 7,
Completing the Urinalysis Strip Full Test
Complete the testing for one strip or continue testing one strip at a
time, until you finish testing all the strips you want to analyze.
To complete the urinalysis strip Full Test, perform the following steps:
1. Remove the used urinalysis strip from the test table, and dispose of
it according to your standard laboratory procedures.
2. Wipe the table insert, if necessary (see Section 4,
Weekly Cleaning of the Test Table and Test Table Insert
3. Report the results to a laboratory supervisor or physician.
4. Select
Done
to complete the test and return to the
Select Ready
screen.
5. Select
Done
to return the
Strip Test Prepare
screen.
You are ready to start the next test. If you completed your testing,
select
Back
to return to the
Select
screen.
Performing an hCG Cassette Full Test
With an hCG cassette full test, you can enter an Operator Name,
Patient Name, and Patient ID. When you place a cassette on the test
table, the analyzer calibrates and starts to perform the analysis.
Perform the following procedures to test a cassette.
Entering Operator and Patient Information
Enter or select an operator name, patient name, and patient ID.
BIOHAZARD
Wear personal protective equipment. Use universal
precautions. For recommended precautions when you work
with biohazardous materials, see
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...