
72
CLINITEK Analyzer Operator’s Guide
Troubl
eshooting
E59
Inverted strip
positioned on the
test table
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
Section 2,
E60
Tilted strip
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
E61
Dry strip
Repeat the test with a fresh strip and
ensure that the strip has been in
contact with the sample (see Section
2,
E62
Light Ingress
Too much light is reflecting on the
analyzer. Move the analyzer to a
location with lower lighting.
Contact your local technical support
provider or distributor.
E63
Failure to find end
of strip
Repeat the test with a fresh strip and
ensure that the strip is correctly
positioned on the test table (see
Section 2,
Error
Code
Error Message
Action
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...