CLINITEK Analyzer Operator’s Guide
35
Op
erations
Performing an hCG Cassette Quick Test
With an hCG cassette Quick Test, when you place the hCG cassette on
the test table, the analyzer calibrates and starts to perform the
analysis. Perform the following tasks to run an hCG cassette Quick
Test.
Preparing a Cassette Quick Test
Note
For more information about the use and storage of test
cassettes, see the Clinitest hCG cassette test instructions for use insert.
To prepare a cassette Quick Test, perform the following steps:
1. On the the
Select Ready
screen, select
Cassette Test
.
2. If configured, enter the cassette lot number and expiration date:
•
To use the last cassette number and begin the test, select
Use
Last Lot
.
•
To enter new cassette data, select
Enter new lot and
expiration
. Enter the cassette lot number and select
Enter
.
Use the arrow keys to enter the cassette expiration date and
select
Enter
.
3. On the
Test Type
screen, select
Clinitest hCG cassette
.
The
Prepare Test
screen displays.
Note
To display the cassette testing steps on the screen, select
Help
.
4. Position the test table insert in the test table for a cassette test.
BIOHAZARD
Wear personal protective equipment. Use universal
precautions. For recommended precautions when you work
with biohazardous materials, see
CAUTION
Bring the test cassette and urine sample to room temperature
20° to 30°C (68° to 86°F) before you run a test. The wrong
temperature could cause inaccurate test results.
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...