46
CLINITEK Analyzer Operator’s Guide
Op
er
at
ions
4. Enter the patient ID (maximum of 13 characters) on the
Enter
Patient ID
screen, and select
Enter
.
Preparing a Cassette Full Test
Note
For more information about the use and storage of test
cassettes, see the Clinitest hCG cassette test instructions for use insert.
To prepare a cassette Full Test, perform the following steps:
1. On the the
Select Ready
screen, select
Cassette Test
.
2. If configured, enter the cassette lot number and expiration date:
•
To use the last cassette number and begin the test, select
Use
Last Lot
.
•
To enter new cassette data, perform the following steps:
a. Select
Enter new lot and expiration
.
b. Enter the cassette lot number and select
Enter
.
c. Use the arrow keys to enter the cassette expiration date and
select
Enter
.
3. On the
Test Type
screen, select
Clinitest hCG cassette
.
The
Prepare Test
screen displays.
Note
To display the cassette testing steps on the screen, select
Help
.
4. Position the test table insert in the test table for a cassette test.
5. Remove the test cassette from the foil package and place the
cassette on the test table.
Running a Cassette Full Test
To run the test, you have 8 seconds to perform the following 2 steps:
Note
After you select
START
, you have 8 seconds to draw the
urine sample into the pipette and add the urine sample into the
well on the cassette.
CAUTION
Bring the test cassette and urine sample to room temperature
20° to 30°C (68° to 86°F) before you run a test. The wrong
temperature could cause inaccurate test results.
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...