126
CLINITEK Analyzer Operator’s Guide
System C
onf
iguration
Running Diagnostics
You can run the following diagnostics on the analyzer:
•
Display
•
Touch Screen
•
Printer
•
Test Table
•
Light Source
•
Electronics
•
Check cassette
To run the diagnostics, perform the following steps:
1. On the
Select Ready
screen, select
Instrument Set Up
.
2. On the
Choose Settings
screen, select
Instrument Settings
.
3. On the
Instrument Settings
screen, select
Diagnostics
.
4. On the
Select Diagnostics Test
screen, select a diagnostic test
5. Select
Select
.
6. Read the onscreen instructions.
7. Select
Run Test
.
8. Select
Done
twice.
Viewing the System Information
You can view the following system information:
•
Serial number
•
Software version
CAUTION
Do not run the Check Cassette diagnostics on your own. Run
the Check Cassette diagnostic tests only when your local
technical support provider or distributor asks you to do so.
The representative will lead you through the test procedure.
For local technical support providers and distributors, see
Appendix B, Support Information
.