CLINITEK Analyzer Operator’s Guide
41
Op
erations
Note
To display the strip testing steps on the screen, select
Help
.
2. Dip the reagent strip in the urine sample and wet all the pads.
The ID band allows auto-strip identification to ensure that the
analyzer reports the correct strip configuration when you perform
a urinalysis test.
Note
Be sure to use the proper dipping technique.
3. Immediately remove the strip from the urine.
4. Drag the edge of the strip against the side of the sample container
as you remove it.
5. Blot the edge of the strip on a paper towel to remove the excess
urine.
6. Place the reagent strip in the test table channel with the test pads
facing up.
7. Slide or push the strip to the end of the channel. Do not touch the
pads on the strip.
After the 8-second countdown ends, the analyzer pulls in the test
table and strip, and then calibrates.
Note
Each time you run a test, the analyzer calibrates.
After the calibration finishes, the analyzer starts analyzing the
strip, and the
Analyzing
screen displays.
Selecting the Appearance of the Urine Sample
While the analyzer analyzes the strip, a
Select Appearance
screen
displays.
To select the appearance of the urine sample, perform the following
steps:
CAUTION
Do not push or pull the test table because the calibration
might fail or the movement might cause table positioning
errors.
Do not move or bump the table while the analyzer calibrates.
The calibration might fail.
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...