CLINITEK Analyzer Operator’s Guide
125
System C
onf
iguration
•
Then, you hear repeated beeps at a higher tone for up to 2
minutes.
When the installation finishes, the analyzer performs the following
operations:
•
The screen displays the message,
Performing a System
Diagnostic Test
.
•
The test table retracts and extends.
•
The
Software Update
screen displays a message that the
software was successfully installed.
6. Select
Done
.
If you upgrade from software Version 1.x, the system displays the
message
E27, Set Up Failure
. Clear the error message by
continuing with the steps in the next section on how to complete
the software upgrade.
Note
The E27 message indicates that a significant change was
made to the system database and occurs with a successful
software upgrade from software Version 1.x.
To complete the software upgrade, perform the following steps:
1. Press the on/off power button until the analyzer powers off.
The test table retracts.
2. Remove the memory card from the memory card slot.
3. Close the printer cover.
4. Power on the analyzer.
CAUTION
Do not leave the memory card in the slot after you finish the
upgrade. If you do, the system deletes all sample results and
performs an unnecessary upgrade each time you power on
the analyzer.
Summary of Contents for CLINITEK Status+
Page 28: ...28 CLINITEK Status Analyzer Operator s Guide ...
Page 50: ...50 CLINITEK Status Analyzer Operator s Guide Operations ...
Page 56: ...56 CLINITEK Status Analyzer Operator s Guide Calibration QC ...
Page 84: ...84 CLINITEK Status Analyzer Operator s Guide File Management ...
Page 128: ...128 CLINITEK Status Analyzer Operator s Guide System Configuration ...