Examination
Fluoroscopy
AXIOM Iconos R200
AXD3-340.620.13.01.02
11
/ 52
Fluoroscopy time limit
Fluoroscopy is switched off automatically and can no longer be switched on after
a fluoroscopy time of 10 minutes
1
.
To continue working
◆
let the fluoroscopy switch go and press it once again.
Fluoroscopy warning signal
An acoustic signal sounds after a fluoroscopy time of maximum 5 minutes
2
.
◆
The signal disappears when you press the key on the Fluorospot Compact.
The time for the acoustic signal can be reset to zero at any time.
Automatic format collimation in fluoroscopy
Automatic format collimation is also switched on automatically when fluoroscopy
is released.
I.e. the fluoroscopic field is never larger than the selected image intensifier for-
mat.
You can limit the fluoroscopic field to the required size with the corresponding
operating elements within this area.
Fluoroscopy programs
All parameters necessary for fluoroscopy are stored in programs.
The following programs can be stored:
– Standard fluoroscopy
– Supervision
– Pulsed fluoroscopy (Carevision)
– Roadmap standard fluoroscopy
– Roadmap Supervision
– Roadmap pulsed fluoroscopy
➩
Refer to the Fluorospot Compact Operator Manual.
1
Can be changed on request by SIEMENS Uptime Service, under compliance with the national
regulations.
2
Can be configured from 1 to 5 minutes by the SIEMENS Uptime Service
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...