System Overview
Protective Measures
14
/ 44
AXD3-340.620.13.01.02
Operator Manual
Stray radiation in the main operating area
according to DIN EN 60601-1-3
Tolerance of the air kerma measurements ± 5%
Characteristic A and C: continuous fluoroscopy 63kV, 0.8mA (antiisowatt)
Characteristic B and D: continuous fluoroscopy 110kV, 3mA
Patient table horizontal
Patient table vertical
0
1
2
3
4
5
6
0
50
100
150
200
cm
mGy/h
Height above floor
Air kerma
A
B
0
1
2
0
50
100
150
200
cm
mGy/h
Height above floor
Air kerma
C
D
Summary of Contents for AXIOM Iconos R200
Page 2: ......
Page 12: ...System Overview Table of Contents 4 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 18: ...System Overview System Description 10 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 32: ...System Overview Protective Measures 24 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 52: ...System Overview Operating and Display Elements 44 44 AXD3 340 620 13 01 02 Operator Manual ...
Page 56: ...System Operation Table of Contents 4 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 78: ...System Operation System Settings 26 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 84: ...System Operation Setting the Image Geometry 32 32 AXD3 340 620 13 01 02 Operator Manual ...
Page 132: ...Examination Digital Radiography 48 52 AXD3 340 620 13 01 02 Operator Manual ...
Page 172: ...POLYDOROS SX 65 80 Integrated Control Console 36 36 AXD3 340 620 13 01 02 Operator Manual ...
Page 184: ...Accessories Standard Accessories 12 54 AXD3 340 620 13 01 02 Operator Manual ...
Page 228: ...Technical Description Table of Contents Operator Manual 2 of 12 AXD3 340 620 13 01 02 ...